x-ray units



Devices of the T-300 T-200 T-160 x-ray therapy unit family can be used for superficial and
orthovoltage treatments in a range from 15 kV to 300 kV over the whole family range, designed to meet today’s clinical requirements. The high tension / tube current outputs of the tube can be combined with seven clinical filters up to twenty HVL combinations that gives the user the widest range of clinical options
The therapy unit is controlled by a patient verification system that has been designed specifically for the x-ray therapy unit family of superficial and orthovoltage x-ray therapy systems. It is a included part of the system. The intuitive interface enables quick and easy access to all patient and treatment history data.
Treatment fields can be created quickly and simply with the help of many input assistance reducing the time required for data entry.
A record of each treatment is stored under the patient’s unique identification.
Instant access to this data enables the operator to easily review a patient’s entire
course of treatment on screen.
The patient records are stored in a dedicated central database, which can be
archived or exported in a lot of common formats for data analysis. A separate report generator allows a variety of report printing options including individual treatment or patient records and user customizable system logs.
The System ioRT-50 completes the range of equipment for conventional radiotherapy by a further system for superficial treatment.

The system T-105 is intended for use for the superficial x-ray therapy. It is offered in two different energy ranges. On the one hand are devices with the maximum tube voltage of 100 kV available . On the other hand, the version offered up to max. 50 kV is possible to be used as directly patient-oriented.


 
brochure x-ray therapy unit T-105



max. high tension [kV] 100
max. tube current [mA] 15
max. power [W] 1.200
two circuit cooler water - water
number of selectabale energies
(combined of filter, [mA], [kV])
4
20 (Option : Verifikationssystem)
max. number of applicators 14
focal - scin distances [cm] 15 ... 30
mains 1 ~ PEN , 230 V AC, 16 AT
   

The T-160 is a standalone solution, specifically designed for sites with limited space.
The available energy allows the application from superficial x-ray treatment to the orthovoltage therapy.

As an option the system can be expanded to a networking verification & record system ...




 
brochure x-ray therapy unit T-160




max. high tension [kV] 150
max. tube current [mA] 20
max. power [W] 3000
two circuit cooler water - water
number of selectabale energies
(combined of filter, [mA], [kV])
20
max. number of applicators 15
focal - scin distances [cm] 20 ... 50
mains 1 ~ PEN , 230 V AC, 35 AT
   

With its broad energy spectrum, the T-200 system allows the use from the field of superficial x-ray treatment to the orthovoltage therapy. Integral part is a networking verification & record system with powerful functions for treatment planning and documentation - of course, completely paperless. Convenient x-ray-tube-stand solutions, whether 3 - D ceiling mount ...






 

... or ground-based column stand, are to predestine the T-200 as the standard solution for hospitals and practices. A variety of options are to complete the system to an effective therapy unit.






 
brochure x-ray therapy unit T-200



max. high tension [kV] 200
max. tube current [mA] 20
max. power [W] 3000
two circuit cooler water - water
number of selectabale energies
(combined of filter, [mA], [kV])
20
max. number of applicators 15
focal - scin distances [cm] 20 ... 50
mains 1 ~ PEN , 230 V AC, 35 AT
   

The powerful 300 kV unit provides high dose rates for the orthovoltage-deep-therapy. The choice of "soft" spectra for the superficial x-ray treatment is also possible. The T-300 completes effectively the canon of the clinical practice of procedures in radiotherapy. Integral part of a powerful networked verification & record system for treatment planning and documentation - of course, completely paperless. With the easy to use 3 - D Ceiling Mount also in large treatment rooms, a variety of settings are possible. A variety of options are to complete the system to an effective therapy unit.







 
brochure x-ray therapy unit T-300



max. high tension [kV] 300
max. tube current [mA] 20
max. power [W] 3000
two circuit cooler oil - water
number of selectabale energies
(combined of filter, [mA], [kV])
20
max. number of applicators 15
focal - scin distances [cm] 20 ... 50
mains 1 ~ PEN , 230 V AC, 35 AT
   

The ioRT-50 ist a special x-ray therapy system using a transmission target tube (Chaoul) for the treatment of superficial cancers of the skin or in body cavities or temporarily accessible cancer debridements.







 
brochure x-ray therapy unit ioRT-50



max. high tension [kV] 70
max. tube current [mA] 7
max. power [W] 500
two circuit cooler water - water
number of selectabale energies
(combined of filter, [mA], [kV])
1
max. number of applicators 255
focal - scin distances [mm] 15 ... 30
mains 1 ~ PEN , 230 V AC, 10 AT